FTIR spectrophotometer, Nicolet Instruments Magna-IR 550

Nicolet Magna-IR 550 spectrometer is capable of up to 0.125 cm-1 resolution. The spectrometer is designed for mid-spectral range coverage. Spectra can be collected and summed at a rate of up to 20 scans per second. The bench is controlled via Nicolets’s Omnic software with a full diagnostics analysis system. The source is a tungsten-halogen source. A variable-diameter aperture is used to achieve the maximum spectral resolution. The detector/beamsplitter combination is a DTGS-KBr providing a spectral range of 4000-400 cm-1.
The bench is continuously purged with dry air reduced CO and moister from a Balston Filter Products 75-20 dry air system. A Harrick Reflection Adsorption attachment allows analysis of monolayer films (RA-FT-IR) and a Harrick Attenuated Total Reflectance attachment allows multiple pass surface analysis at varied sampling depths using a range of angled Ge crystals.

Refurbished, Infrared Spectrophotometer, Perkin Elemr 1310

This is a Perkin Elmer 1310 Infrared Spectrophotometer. Automatic-recording, double-beam, optical null instrument which features microprocessor-controlled abscissa functions, a DC wedge/pen servo mechanism, a four-digit frequency readout, and a touch-control panel. The monochromator, photometer and electronics compartment houses the grating monochromator, the detector, and the associated electronics. The source compartment contains the source, the power supply and the fuses.

  • Abscissa Range 4000 - 600 cm-1
  • Abscissa Accuracy: 4000 - 2000 cm-1: ±6 cm-1 ; 2000 - 600 cm-1: ±3 cm-1
  • Abscissa Expansion x0.5, x1, x5
  • Ordinate Accuracy ±2 T
  • Ordinate Repeatability within 0.75 T100
  • Line Flatness ±3 T
  • Signal/Noise Better than 100:1
  • Resolution (Narrow Slit Program) 3 cm-1 at 1000 cm-15 cm-1 at 3000 cm-1
  • Stray Light < 1T (4000 - 450 cm-1)< 2T (450 - 250 cm-1)< 3T (250 - 200 cm-1)
  • Time Drive 0.5, 1 and 5 cm/min